![]() ; Francis, Olivier ![]() in Metrologia (2017), 54(1A), 07012 In the framework of the regional EURAMET.M.G-K2 comparison of absolute gravimeters, 17 gravimeters were compared in November 2015. Four gravimeters were from different NMIs and DIs, they were used to link ... [more ▼] In the framework of the regional EURAMET.M.G-K2 comparison of absolute gravimeters, 17 gravimeters were compared in November 2015. Four gravimeters were from different NMIs and DIs, they were used to link the regional comparison to the CCM.G.K2 by means of linking converter. Combined least-squares adjustments with weighted constraint was used to determine KCRV. Several pilot solutions are presented and compared with the official solution to demonstrate influences of different approaches (e.g. definition of weights and the constraint) on results of the adjustment. In case of the official solution, all the gravimeters are in equivalence with declared uncertainties. == Main text To reach the main text of this paper, click on Final Report [http://www.bipm.org/utils/common/pdf/final_reports/M/G-K2/EURAMET.M.G-K2.pdf] . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ [http://kcdb.bipm.org/] . The final report has been peer-reviewed and approved for publication by the CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). [less ▲] Detailed reference viewed: 360 (30 UL)![]() Francis, Olivier ![]() in Metrologia (2015), 52(1A), 07009 In November 2013 an International Key Comparison, CCM.G-K2, was organized in the Underground Laboratory for Geodynamics in Walferdange. The comparison has assembled 25 participants coming from 19 ... [more ▼] In November 2013 an International Key Comparison, CCM.G-K2, was organized in the Underground Laboratory for Geodynamics in Walferdange. The comparison has assembled 25 participants coming from 19 countries and four different continents. The comparison was divided into two parts: the key comparison that included 10 NMIs or DIs, and the pilot study including all participants. The global result given by the pilot study confirms that all instruments are absolutely coherent to each other. The results obtained for the key comparison confirm a good agreement between the NMI instruments. Main text. To reach the main text of this paper, click on Final Report [http://www.bipm.org/utils/common/pdf/final_reports/M/G-K2/CCM.G-K2.pdf] . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ [http://kcdb.bipm.org/] . The final report has been peer-reviewed and approved for publication by CCM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA). [less ▲] Detailed reference viewed: 505 (47 UL) |
||